Differently zoomed images of same device requires individual calibration for accurate measurement of object in an images.how to manage a single calibration on such images ?

1 回表示 (過去 30 日間)
ultrasound images of same device differently zoomed and saved for further processing

採用された回答

Image Analyst
Image Analyst 2013 年 10 月 18 日
You can't. Different zoom levels will have a different calibration scale factor and there is no way around that if you want accurate lengths and areas out.
  2 件のコメント
vaishali
vaishali 2013 年 10 月 18 日
Resize/zooming (at same level) all the images and then calibrating, processing & taking measurement.then possible???
Image Analyst
Image Analyst 2013 年 10 月 18 日
編集済み: Image Analyst 2013 年 10 月 18 日
Yes, of course. If you recalibrate after you change the field of view then the measurements will be accurate.