Differently zoomed images of same device requires individual calibration for accurate measurement of object in an images.how to manage a single calibration on such images ?
1 回表示 (過去 30 日間)
古いコメントを表示
ultrasound images of same device differently zoomed and saved for further processing
0 件のコメント
採用された回答
Image Analyst
2013 年 10 月 18 日
You can't. Different zoom levels will have a different calibration scale factor and there is no way around that if you want accurate lengths and areas out.
2 件のコメント
Image Analyst
2013 年 10 月 18 日
編集済み: Image Analyst
2013 年 10 月 18 日
Yes, of course. If you recalibrate after you change the field of view then the measurements will be accurate.
その他の回答 (0 件)