DAQ with ENET-9163 and NI 9237 National Instruments

Hello,
I can't open a channel of the NI 9237 Device in Matlab using R2011a is it a device support problem or is there any solution? That's the device info i'm getting:
>> get(deviceInfo)
IsSimulated: false
Vendor: National Instruments
ID: 'ENET-14E1636'
Model: 'ENET-9237'
Subsystems: empty
Description: 'National Instruments ENET-9237'
RecognizedDevice: false
and when i want to open a analoginput:
s.addAnalogInputChannel('ENET-14E1636', 'ai0', 'Voltage');
??? The requested subsystem 'AnalogInput' does not exist on this device.
Thanks for any help!

回答 (2 件)

Manisha
Manisha 2012 年 2 月 21 日

0 投票

Hi Dirk,
Support for NI 9237 was added in R2011b. NI 9237 supports only bridge measurements, so if you use R2011b , you should be able to use this device as
s = daq.createSession('ni');
s.addAnalogInputChannel('Dev5', 0, 'Bridge');
Look at this demo for more information on bridge channels.
Hope that helps,
Manisha
Vinh Dang
Vinh Dang 2015 年 3 月 5 日
編集済み: Vinh Dang 2015 年 3 月 5 日

0 投票

Hello, In my application, I have a cDAQ configuration with AI modules 9234 (for accelerometers), 9237 (for pressure and load cell) and 9215 (for proximity and temperature). All modules are put in one chassis cDAQs-9178. What is the valid sampling rate for this?
Thanks for your help!