Finding high and low frequency noise....

1 回表示 (過去 30 日間)
Tinkul
Tinkul 2014 年 5 月 3 日
回答済み: Star Strider 2014 年 5 月 3 日
Is power line interference high frequency noise in ecg and base line wandering low frequency noise in ecg? Is both the noise present in EMG and EEG signal?
  3 件のコメント
Tinkul
Tinkul 2014 年 5 月 3 日
it is biomedical signal processing question..
Image Analyst
Image Analyst 2014 年 5 月 3 日
OK, but do you have a MATLAB question for this MATLAB forum? If you do, attach your m-file. http://www.mathworks.com/matlabcentral/answers/6200-tutorial-how-to-ask-a-question-on-answers-and-get-a-fast-answer

サインインしてコメントする。

採用された回答

Star Strider
Star Strider 2014 年 5 月 3 日
Powerline noise can appear in poorly-instrumented EMG and EEG signals. The usual way to deal with this in hardware is to have a good subject reference ground, no ground faults (patient ground and instrument ground are the same), and to use shielded cables.
You can get rid of the 50-60 Hz noise in the recorded signal using a digital notch filter. Baseline wandering is usually dealt with by using an appropriately-designed bandpass filter.