- You will likely have to modify the SCPI commands to match your scope model
- If you modify the code, remember that the 2 main reasons for the data not coming back correctly are incorrect ByteOrder (you're reading bigEndian while the scope is sending back littleEndian data, for example) or incorrect precision (the scope is sending back uint16 data and you are reading int16 data, for example).
How to get correct data from OSC Agilent?
2 ビュー (過去 30 日間)
古いコメントを表示
Dear Support team,
I am trying to get data from Agilent OSC DSOX96204Q using "binblockread()" function, but the retrieved data is not correct. The OSC is always displays streaming continuous data and I can not acquire correct data from OSC. How can we get correct data?
+ By acquiring instantaneous signal using a loop?
+ By acquiring signal in many times with interval for each time? If this is the case, what is the parameter on the instrument device need to setup for the acquiring signal?
Thank you
Hase.